Switzerland and the Principality of Liechtenstein
Radio Interface Regulation

784.101.21 / RIR10061)
Wireless applications in healthcare

RIR1006-12
ULP-WMCE
430 - 440 MHz
Ultra Low Power Wireless Medical Capsule Endoscopy (ULP-WMCE)
Edition 1.0 / 01.05.2020

Nr Parameter2)DescriptionComments
1Radiocommunication service--
2ApplicationULP-WMCEUltra Low Power Wireless Medical Capsule Endoscopy (ULP-WMCE) application, which includes Capsule Camera (CCam) acting as transmitter and associated Data Recorder (DR) receiver devices, covered by the applicable harmonised standard.
3Frequency band430 - 440 MHz-
4Channelling--
5Modulation / Occupied bandwidth- / OBW: Max. 10MHz.-
6Direction / Separation--
7Transmit power / Power density-50 dBm/100kHz max e.r.p. density but not exceeding a total power of -40 dBm/10MHzBoth limits are intended for measurement outside of the patient's body.
8Channel access and occupation rules--
9Authorisation regimeSwitzerland: License exempt.
Liechtenstein: General authorisation.
Operation on NIB / NPB. (Non interference basis / non protection basis).
10Additional essential requirements--
11Frequency planning assumptionsEN 303 520-
12Planned changes--
13ReferenceEN 303 520 /-/ ERC/REC 70-03, Annex 13.Effective use of spectrum / ECC Decisions / ECC Recommendations.
14Notification numberG/TBT/N/CHE/242
2020/9501/CH
-
15RemarksEN 301 489-3-

Normative part: 1 to 11
Informative part: Nr 12 to 15

1) RIR for similar and/or other applications are available under: https://www.bakom.admin.ch/bakom/en/home/frequenzen-antennen/nationaler-frequenzzuweisungsplan/schnittstellen-anforderungen.html
2) For explanations and legal status, please refer to basis document 784.101.21 / RIR0000