Switzerland and the Principality of Liechtenstein
Radio Interface Regulation

784.101.21 / RIR10061)
Wireless applications in healthcare

RIR1006-01
ULP-AMI
9 - 315 kHz
30 dBuA/m / 10% DC / ULP-AMI
Edition 3.0 / 01.08.2018

Nr Parameter2)DescriptionComments
1Radiocommunication service--
2ApplicationULP-AMIUltra Low Power Active Medical Implants (ULP-AMI).
Covers the radio part of active implantable medical devices and their peripherals.
3Frequency band9 - 315 kHz-
4Channelling--
5Modulation / Occupied bandwidth--
6Direction / Separation--
7Transmit power / Power densityMax. 30 dBuA/m at 10m-
8Channel access and occupation rulesDuty Cycle max. 10 %-
9Authorisation regimeSwitzerland: License exempt.
Liechtenstein: General authorisation.
Operation on NIB / NPB. (Non interference basis / non protection basis).
10Additional essential requirements--
11Frequency planning assumptionsEN 302 195 or EN 300 330-
12Planned changes--
13ReferenceEN 302 195 or EN 300 330 /-/ ERC/REC 70-03, Annex 12.Effective use of spectrum / ECC Decisions / ECC Recommendations.
14Notification numberG/TBT/N/CHE/102
2008/9503/CH
-
15RemarksEN 301 489-31-

Normative part: 1 to 11
Informative part: Nr 12 to 15

1) RIR for similar and/or other applications are available under: https://www.bakom.admin.ch/bakom/en/home/frequenzen-antennen/nationaler-frequenzzuweisungsplan/schnittstellen-anforderungen.html
2) For explanations and legal status, please refer to basis document 784.101.21 / RIR0000